Unique push fit elements ensure quick and simple maintenance.
Key specs:
Flow Rate
10 to 2549 Nm3/h
Connection
⅛” to 3” BSPP
Filtration
0.01 micron
Product Documentation
- Brochures
- Spare Parts Guide
- User Guides
Features
“Dead Stop” bowl feature provides full closure without over tightening.
Housing manufactured from cast aluminum alloy providing enhanced strength and feature an E-Coat™ finish for optimum internal and external corrosion protection.
Specifications of Our Models
Medical Sterile Filter Model |
Replacement Element | Inlet & Outlet | Rated Flow at 7 barg |
---|---|---|---|
Part No. | BSPP | Nm3/h | |
GMSB 0006 | GE 0015 MS | ⅛” | 10 |
GMSB 0015 | GE 0015 MS | ¼” | 26 |
GMSB 0025 | GE 0032 MS | ¼” | 43 |
GMSB 0032 | GE 0032 MS | ⅜” | 54 |
GMSB 0050 | GE 0050 MS | ½” | 85 |
GMSB 0070 | GE 0105 MS | ½” | 119 |
GMSB 0085 | GE 0105 MS | ¾” | 144 |
GMSB 0105 | GE 0105 MS | 1” | 178 |
GMSB 0125 | GE 0175 MS | ¾” | 212 |
GMSB 0175 | GE 0175 MS | 1” | 297 |
GMSB 0280 | GE 0325 MS | 1 ¼” | 476 |
GMSB 0325 | GE 0325 MS | 1 ½” | 552 |
GMSB 0450 | GE 0450 MS | 2” | 765 |
GMSB 0700 | GE 0700 MS | 2” | 1189 |
GMSB 0850 | GE 0900 MS | 2 ½” | 1444 |
GMSB 0900 | GE 0900 MS | 3” | 1529 |
GMSB 1250 | GE 1250 MS | 3” | 2124 |
GMSB 1500 | GE 1500 MS | 3” | 2549 |
Specifications |
GMSB 0006 to GMSB 0015 |
GMSB 0025 to GMSB 0050 |
GMSB 0070 to GMSB 1500 |
---|---|---|---|
Design operating pressure range |
1.5 to 20.7 barg |
1.5 to 20.7 barg |
1.5 to 20.7 barg |
Design operating temperature range | 2 to 120°C | 2 to 120°C | 2 to 120°C |
Condensate drain |
GFMDK 0050 | GFMDK 0050 |
GFMDK 1500 |
Differential pressure indicator/gauge |
- | indicator GFDP 1500 |
gauge GFDP 1500 |
Element Performance |
MS |
---|---|
DOP efficiency |
>99.9999% |
Particle removal |
0.01 micron |
Recommended operating temperature range |
2 to 100°C |
Design operating temperature range |
2 to 120°C |
Maximum autoclave temperature | 133°C |
Pressure drop - clean | 0.1 bar |
Pressure drop - recommended replacement | 0.4bar |
Endcap material | stainless steel |
Flow direction through element | outside to inside |
Maximum steam sterilisation cycles | 100 |
Maximum element life | 6 months or 1000 hours |
(1) This information is for general information purposes only. For additional technical specifications, refer to the product brochure, technical data sheets, and/or user guides. Technical specifications are subject to change.
Frequently Asked Questions
A medical sterile compressed air filter is designed to ensure that the compressed air used in medical environments is free from contaminants such as live bacteria and other harmful sub-micron particles. It is used for critical air quality, such as dental, medical, and veterinary applications.
To protect equipment from airborne contaminants, sterile compressed air filters are essential in medical settings. Contaminated air can lead to infections, compromise sensitive procedures, or damage medical equipment. The filter ensures the air is free from harmful sub-micron particles and live bacteria.
These filters utilise a multi-stage filtration process to eliminate contaminants. They typically consist of layers designed to capture sub-micron particles and live bacteria.
Standard compressed air filters are inadequate for medical applications as they fail to meet strict air purity and sterilisation standards. Medical sterile compressed air filters are specially engineered to guarantee safe air quality for healthcare purposes.
- Assess the Application: Understand where and how the compressed air will be used (e.g., operating rooms, ventilators, laboratories).
- Identify Filtration Requirements: Determine the level of sterility required and whether moisture removal is necessary.
- Review Compliance Standards: Ensure the filter meets the necessary regulations.
- Check Performance Specifications: Evaluate flow rate, pressure drop, and the capacity of the filter to maintain system performance.
- Consider Maintenance Needs: Factor in ease of maintenance, replacement schedules, and filter longevity.
Our medical sterile filters are designed to remove various contaminants, including live bacteria and sub-micron particles.
The GMSB range has a micron rating of 0.01 or >99.9999%, indicating the size of particles the filter can capture.
Yes, the GMSB filter elements are compatible with autoclave sterilisation and are guaranteed for a minimum of 100 sterilisations at 120°C.
The maximum design operating pressure is 20.7 barg. The pressure rating indicates the maximum pressure the filter can withstand without collapsing or damaging. Ensure the filter is rated to handle your compressed air system pressures to maintain efficiency and avoid potential system failure.
Our filters are designed to meet the highest industry standards, including ISO 7396-1, HTM 02-01, AS2896, and NFPA 99, making them the go-to choice for dental, medical, and veterinary applications.
The nano GMSB filter housings are made from die-cast aluminium alloy and feature a durable powder-coated exterior and an internal and external E-Coat™ finish for superior corrosion resistance. The filter element is constructed of a borosilicate microfiber with stainless steel endcaps.
Filters require regular maintenance to function correctly—contact nano for the proper maintenance schedule.
The filter element should be replaced every 6 months or 1000 hours.
The cost varies depending on the size required for your compressed air system.
18 months from the date of shipment (from the factory) or 12 months from the date of installation/start-up, whichever occurs first • Coverage: 100% parts only